Function Group
SYSTEM
Menu Map
Description
This
sets the standard starting configuration of the swept
signal generator when the
key is pressed, as set by the
manufacturer. An asterisk next to the key label indicates that this
feature is active. The following is a description of the configuration.
n
Start sweep at the minimum specified frequency.
Stop sweep at the maximum specified frequency.
n
Power level set at 0
n
Sweep time set to auto.
n
CONT sweep.
n
Sweep mode ramp.
n
ALC leveling point internal.
n
ALC leveling mode normal.
n
Markers set to activate at the center frequency of the sweep.
n
All function values stored in memory registers 1 through 9 remain
in their previous states.
n
The checksum of the calibration data is calculated, and if an error
is detected, the calibration data in protected memory is used. If
the checksum of the protected data is not correct, then default
values are used an error message (EEROM FAILED, LOST CAL)
is displayed.
Programming Codes
SCPI:
SYSTem:PRESet:TYPE
Analyzer:
IP, which is the same as
See Also
Preset Mode User
“Changing the Preset Parameters” in Chapter 1
P-10 Operating and Programming Reference
Содержание 8360 B Series
Страница 231: ...HP IB Henu See Also CONNECTORS HP IB Getting Started Programming in Chapter I H 2 Operating and Programming Reference ...
Страница 363: ...NF N 2 w z P w D I oa 0 i B 6 7 On s5 n 20 a g 9 i T2 ...
Страница 406: ...2b Menu Maps Menu Maps 2b 1 ...
Страница 407: ...t t m ALC Bandwidth Select Module Sense Low High Auto AUTO Front Rear None ALC MENU ...
Страница 412: ...rMENU SELECT 1 Fault Fault Info1 Info2 Fault Info3 Clear Fault S E R V I C E M E N U 6 ...
Страница 469: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...