ANALYZER STATUS
REGISTER
Function Group
NONE
Menu Map
NONE
Description
The following is the status register structure of the swept signal
generator when the analyzer programming language is selected. This
status structure is the structurally and syntactically the same as on
the HP
Output Status bytes, is used to read the two
status
bytes from the swept signal generator. The first status byte
concerns the cause of an SRQ (Service Request), while the
second status byte concerns failures and faults, as follows:
STATUS BYTE
Bit #
7
6
5
4
3
2
1
Decimal
126
64
32
16
4
2
1
Value
Function
SRQ on new REQUEST
SRQ on
SRQ on
SRQ on
SRQ on
SRQ on
SRQ on
frequencies
SERVICE
HP-IB or
End of
RF Settled
Changed in
Numeric
Any Front
or sweep
syntax error. Sweep
Extended
Entry
Panel Key
time in
status
Completed
Pressed
effect.
Byte
(HP-IB or
Front
Panel)
EXTENDED STATUS BYTE
Bit #
7
6
5
4
3
2
1
Decimal
128
64
32
6
8
4
2
1
Value
Function
Fault
RF Unleveled Power
RF Unlocked External
Oven
Self Test
Indicator
Failure
y
Cold
Modulation
Failed
On
Reference
Selected
Status Byte 1
Bit 0: SRQ caused by a key closure on the front panel of the swept
signal generator (use the OM code to determine the front panel
status).
Bit 1: SRQ caused by the completion of a numeric entry (use the OA
code to determine the value of the numerical entry).
Bit 2: SRQ caused by a change in the extended status byte (status
byte 2) affected by the RE-coded mask (see the RE code for an
explanation of this masking).
A-20 Operating and Programming Reference
Содержание 8360 B Series
Страница 231: ...HP IB Henu See Also CONNECTORS HP IB Getting Started Programming in Chapter I H 2 Operating and Programming Reference ...
Страница 363: ...NF N 2 w z P w D I oa 0 i B 6 7 On s5 n 20 a g 9 i T2 ...
Страница 406: ...2b Menu Maps Menu Maps 2b 1 ...
Страница 407: ...t t m ALC Bandwidth Select Module Sense Low High Auto AUTO Front Rear None ALC MENU ...
Страница 412: ...rMENU SELECT 1 Fault Fault Info1 Info2 Fault Info3 Clear Fault S E R V I C E M E N U 6 ...
Страница 469: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...