6. On the swept signal generator, set the power level and start/stop
frequency information as desired.
7. Press
Select
Det
. The power meter is now under swept
signal generator control and is performing the sequence of steps
necessary to generate the compensation information.
If an HP-IB error message is displayed verify that the interface
connections are correct. Check the HP-IB address of the power
meter and ensure that it is the same address the swept signal
generator is using (address 13 is assumed). Refer to the menu
map 8, System, for the key sequence necessary to reach
Meter Adrs .
9. When the operation is complete, (a message is displayed)
disconnect the power meter/sensor. The swept signal generator
has stored the compensation information in its memory and is
using it to calibrate the detector’s output voltage relative to
power.
Getting Started Advanced