Externally Leveling
the Swept Signal
Generator
In externally leveled operations, the output power from the swept
signal generator is detected by an external sensor. The output of this
detector is returned to the leveling circuitry, and the output power
is automatically adjusted to keep power constant at the point of
detection.
Leveling with
Figure l-11 illustrates a typical setup for external leveling. When
Detectors/Couplers
externally leveled, the power level feedback is taken from the external
/Splitters
negative detector input rather than the internal detector. This
feedback voltage controls the ALC system to set the desired RF
output. Refer to Figure A-l in Chapter 2 for a block diagram of the
swept signal generator’s ALC circuitry.
SUEPT SIGNRL
GENERATOR
Figure l-l 1. ALC Circuit Externally Leveled
Getting Started Advanced 1-23