Function Group
Menu Map
Description
Programming Codes
See Also
SWEEP
7
This
selects single sweep mode, aborts any sweep in progress
and initiates a single sweep at a rate determined by the sweep time
function.
If you press
in the middle of a continuous sweep, the sweep is
aborted and the swept signal generator retraces to the starting point
but does not start a sweep. Press
a second time to start the
sweep. The amber LED above the
is lit when the function is
on.
INITiate[:IMMediate]
Analyzer: S2
“Continuous, Single, and Manual Sweep Operation” in Chapter 1
“Programming Typical Measurements” in Chapter 1
Function Group
SYSTEM
Menu Map
Description
This
displays the swept signal generator’s programming
language, HP-IB address, and firmware date code.
Programming Codes
SCPI:
*IDN?
Analyzer:
01
See Also
HP-13
Menu,
SCPI COMMAND SUMMARY