Creating a User Flatness Array Automatically, Example 1
In this example, a flatness array containing correction frequencies
from 4 to 10
at 1
intervals is created. An HP 438B power
meter controlled by the swept signal generator through the interface
bus is used to enter the correction data into the flatness array.
For this example, refer to menu map 5, POWER.
1. The equipment setup shown in Figure 1-18 assumes that if the
setup has an external leveling configuration, the steps necessary
to correctly level have been followed. If you have questions about
external leveling, refer to “Externally Leveling the Swept Signal
Generator”.
Setup Power Meter
2. Zero and calibrate the power meter/sensor.
3. Enter the appropriate power sensor calibration factors into the
power meter.
4. Enable the power meter/sensor cal factor array. For operating
information on the HP 437B power, refer to its operating and
service manual.
5. Connect the power sensor to the point where corrected power is
desired.
H P - I B
I
CABLES
I
OTHER I
DEVICES ,
q DDDaa
HP 4378
POUER
I
, OUTPUT PORT
POUER SENSOR
OUT
IN
-
-
-
-
-
-
-
-
-
- - B - m -
u
d
DEVICE
UNDER
TEST
Figure 1-16. Creating a User Flatness Array Automatically
Note
No other devices can be connected to the HP-IB cable.
1-34 Getting Started Advanced
Содержание 8360 B Series
Страница 231: ...HP IB Henu See Also CONNECTORS HP IB Getting Started Programming in Chapter I H 2 Operating and Programming Reference ...
Страница 363: ...NF N 2 w z P w D I oa 0 i B 6 7 On s5 n 20 a g 9 i T2 ...
Страница 406: ...2b Menu Maps Menu Maps 2b 1 ...
Страница 407: ...t t m ALC Bandwidth Select Module Sense Low High Auto AUTO Front Rear None ALC MENU ...
Страница 412: ...rMENU SELECT 1 Fault Fault Info1 Info2 Fault Info3 Clear Fault S E R V I C E M E N U 6 ...
Страница 469: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...