Freq Offset
Function Group
FREQUENCY
Menu Map
Description
This
lets you set a frequency offset value and applies it to
all frequency parameters. The frequency offset ranges between and
including fllO.O
Changing the frequency offset value changes
the display but does not affect the output frequency. Frequency
multiplier and offset are related as shown by the following equation:
Entered value Displayed Frequency (Frequency Generated
x
Multiplier) + Offset value
Programming Codes
See Also
The factory preset value is 0 Hz. An asterisk next to the key label
indicates that this feature is active.
FREQuency:OFFSet
FREQuency:OFFSet:STATe
Analyzer:
SHFB
FREQUENCY
Freq
Mult
Operating and Programming Reference F-19