Baud Rate Selection (
RS-232
)
You can select one of six baud rates for
RS-232
operation. The rate is set
to 9600 baud when the multimeter is shipped from the factory.
The baud rate can be set only from the front-panel.
•
Select one of the following: 300, 600, 1200, 2400, 4800, or
9600
baud
(factory setting).
•
The baud rate selection is stored in
non-volatile
memory, and
does not
change when power has been off or after a remote interface reset.
•
Front-Panel Operation
:
3: BAUD RATE (I/O MENU)
See also “To Set the Baud Rate,” on page 163.
Parity Selection (
RS-232
)
You can select the parity for
RS-232
operation. The multimeter is
configured for even parity with 7 data bits when shipped from the factory.
The parity can be set only from the front-panel.
•
Select one of the following: None (8 data bits),
Even
(7 data bits), or
Odd (7 data bits). When you set the parity, you are indirectly setting
the number of data bits.
•
The parity selection is stored in
non-volatile
memory, and
does not
change when power has been off or after a remote interface reset.
•
Front-Panel Operation
:
4: PARITY (I/O MENU)
See also “To Set the Parity,” on page 164.
3
Chapter 3 Features and Functions
Remote Interface Configuration
93
Содержание 34401A
Страница 12: ...1 Quick Start 1 ...
Страница 26: ...2 Front Panel Menu Operation 2 ...
Страница 50: ...3 Features and Functions 3 ...
Страница 103: ...4 Remote Interface Reference 4 ...
Страница 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Страница 170: ...5 Error Messages 5 ...
Страница 182: ...6 Application Programs 6 ...
Страница 196: ...7 Measurement Tutorial 7 ...
Страница 214: ...8 Specifications 8 ...
Страница 222: ...Product Dimensions TOP Product Dimensions All dimensions are shown in millimeters 8 Chapter 8 Specifications 223 ...