DC Characteristics
DC Characteristics
Accuracy Specifications
±
( % of r % of range ) [ 1 ]
Function
Range [ 3 ]
Test Current or
Burden Voltage
90 Day
23°C
±
5°C
1 Year
23°C
±
5°C
Temperature
Coefficient /°C
0°C – 18°C
28°C – 55°C
DC Voltage
100.0000 mV
1.000000 V
10.00000 V
100.0000 V
1000.000 V
0.0030 + 0.0030
0.0020 + 0.0006
0.0015 + 0.0004
0.0020 + 0.0006
0.0020 + 0.0006
0.0040 + 0.0035
0.0030 + 0.0007
0.0020 + 0.0005
0.0035 + 0.0006
0.0035 + 0.0010
0.0050 + 0.0035
0.0040 + 0.0007
0.0035 + 0.0005
0.0045 + 0.0006
0.0045 + 0.0010
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
Resistance
[ 4 ]
100.0000
Ω
1.000000 k
Ω
10.00000 k
Ω
100.0000 k
Ω
1.000000 M
Ω
10.00000 M
Ω
100.0000 M
Ω
1 mA
1 mA
100
µ
A
10
µ
A
5
µ
A
500 nA
500 nA
|| 10 M
Ω
0.0030 + 0.0030
0.0020 + 0.0005
0.0020 + 0.0005
0.0020 + 0.0005
0.002 + 0.001
0.015 + 0.001
0.300 + 0.010
0.008 + 0.004
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.020 + 0.001
0.800 + 0.010
0.010 + 0.004
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.040 + 0.001
0.800 + 0.010
0.0006 + 0.0005
0.0006 + 0.0001
0.0006 + 0.0001
0.0006 + 0.0001
0.0010 + 0.0002
0.0030 + 0.0004
0.1500 + 0.0002
DC Current
10.00000 mA
100.0000 mA
1.000000 A
3.000000 A
< 0.1 V
< 0.6 V
< 1 V
< 2 V
0.005 + 0.010
0.01 + 0.004
0.05 + 0.006
0.10 + 0.020
0.030 + 0.020
0.030 + 0.005
0.080 + 0.010
0.120 + 0.020
0.050 + 0.020
0.050 + 0.005
0.100 + 0.010
0.120 + 0.020
0.002 + 0.0020
0.002 + 0.0005
0.005 + 0.0010
0.005 + 0.0020
Continuity
1000.0
Ω
1 mA
0.002 + 0.010
0.008 + 0.020
0.010 + 0.020
0.001 + 0.002
Diode Test
1.0000 V
1 mA
0.002 + 0.010
0.008 + 0.020
0.010 + 0.020
0.001 + 0.002
DC:DC
Ratio
100 mV
to
1000 V
( Input Accuracy ) + ( Reference Accuracy )
Input Accuracy = accuracy specification for the HI-LO input signal.
Reference Accuracy = accuracy specification for the HI-LO reference input signal.
Transfer Accuracy ( typical )
(
24 hour %
of range error
)
2
Conditions:
- Within 10 minutes and
±
0.5°C.
-
Within
±
10% of initial value.
- Following a 2-hour warm-up.
- Fixed range between 10% and 100% of full scale.
-
Using
6
1
⁄
2
digit slow resolution ( 100 PLC ).
- Measurements are made using accepted metrology practices.
24 Hour [ 2 ]
23°C
±
1°C
Chapter 8 Specifications
216
Содержание 34401A
Страница 12: ...1 Quick Start 1 ...
Страница 26: ...2 Front Panel Menu Operation 2 ...
Страница 50: ...3 Features and Functions 3 ...
Страница 103: ...4 Remote Interface Reference 4 ...
Страница 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Страница 170: ...5 Error Messages 5 ...
Страница 182: ...6 Application Programs 6 ...
Страница 196: ...7 Measurement Tutorial 7 ...
Страница 214: ...8 Specifications 8 ...
Страница 222: ...Product Dimensions TOP Product Dimensions All dimensions are shown in millimeters 8 Chapter 8 Specifications 223 ...