Calibration
Security
(continued)
To Unsecure for Calibration
You can unsecure the multimeter
for calibration either from the front panel or remote interface.
The multimeter is secured when shipped from the factory, and the
security code is set to “
HP034401
”.
•
Front-Panel Operation
:
1: SECURED (CAL MENU)
If the multimeter is secured, you will see the above command when
you go into the
CAL MENU
. (If you move across the “commands” level
in the menu, you will notice that the “
2: CALIBRATE
” command is
“hidden” if the multimeter is secured.) To unsecure the multimeter,
select the “parameter” level of the
SECURED
command, enter the
security code, then press Menu Enter.
When you go to the “commands” level in the
CAL MENU
again,
you will notice that the multimeter is unsecured. Notice also that the
“
2: CALIBRATE
” command is no longer hidden and you can perform
a calibration.
•
Remote Interface Operation
:
CALibration:SECure:STATe {OFF|ON},<
code
>
To unsecure the multimeter, send the above command with the same
code used to secure. For example,
"CAL:SEC:STAT OFF,HP034401"
∧
000000 CODE
1: UNSECURED
Chapter 3 Features and Functions
Calibration Overview
96
Содержание 34401A
Страница 12: ...1 Quick Start 1 ...
Страница 26: ...2 Front Panel Menu Operation 2 ...
Страница 50: ...3 Features and Functions 3 ...
Страница 103: ...4 Remote Interface Reference 4 ...
Страница 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Страница 170: ...5 Error Messages 5 ...
Страница 182: ...6 Application Programs 6 ...
Страница 196: ...7 Measurement Tutorial 7 ...
Страница 214: ...8 Specifications 8 ...
Страница 222: ...Product Dimensions TOP Product Dimensions All dimensions are shown in millimeters 8 Chapter 8 Specifications 223 ...