Measurement Tutorial
The HP 34401A is capable of making highly accurate measurements.
In order to achieve the greatest accuracy, you must take the necessary
steps to eliminate potential measurement errors. This chapter describes
common errors found in measurements and gives suggestions to help
you avoid these errors.
Thermal
EMF
Errors
Thermoelectric voltages are the most common source of error in
low-level dc voltage measurements. Thermoelectric voltages are
generated when you make circuit connections using dissimilar metals
at different temperatures. Each metal-to-metal junction forms a
thermocouple
, which generates a voltage proportional to the junction
temperature. You should take the necessary precautions to minimize
thermocouple voltages and temperature variations in low-level voltage
measurements. The best connections are formed using copper-to-copper
crimped connections. The table below shows common thermoelectric
voltages for connections between dissimilar metals.
Copper-to-
Copper
Gold
Silver
Brass
Beryllium Copper
Aluminum
Kovar or Alloy 42
Silicon
Copper-Oxide
Cadmium-Tin Solder
Tin-Lead Solder
Approx.
m
V / °C
<0.3
0.5
0.5
3
5
5
40
500
1000
0.2
5
The HP 34401A’s input terminals are copper alloy.
198
Содержание 34401A
Страница 12: ...1 Quick Start 1 ...
Страница 26: ...2 Front Panel Menu Operation 2 ...
Страница 50: ...3 Features and Functions 3 ...
Страница 103: ...4 Remote Interface Reference 4 ...
Страница 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Страница 170: ...5 Error Messages 5 ...
Страница 182: ...6 Application Programs 6 ...
Страница 196: ...7 Measurement Tutorial 7 ...
Страница 214: ...8 Specifications 8 ...
Страница 222: ...Product Dimensions TOP Product Dimensions All dimensions are shown in millimeters 8 Chapter 8 Specifications 223 ...