Voltmeter Complete Terminal
The rear-panel
VM Comp
(voltmeter complete) terminal provides a
low-true pulse after the completion of each measurement. Voltmeter
complete and external trigger (
see below
) implement a standard hardware
handshake sequence between measurement and switching devices.
External Trigger Terminal
You can trigger the multimeter by applying a low-true pulse to the
rear-panel
Ext Trig
(external trigger) terminal. To use this terminal
from the remote interface, you must select the external trigger source
(
TRIGger:SOURce EXTernal
).
You can use a simple switch to generate an external trigger using the
Ext Trig
input as shown below.
5 V
0 V
Approximately
2
µ
s
Output
5 V
0 V
>1
µ
s
Input
Ext Trig
3
Chapter 3 Features and Functions
Triggering
83
Содержание 34401A
Страница 12: ...1 Quick Start 1 ...
Страница 26: ...2 Front Panel Menu Operation 2 ...
Страница 50: ...3 Features and Functions 3 ...
Страница 103: ...4 Remote Interface Reference 4 ...
Страница 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Страница 170: ...5 Error Messages 5 ...
Страница 182: ...6 Application Programs 6 ...
Страница 196: ...7 Measurement Tutorial 7 ...
Страница 214: ...8 Specifications 8 ...
Страница 222: ...Product Dimensions TOP Product Dimensions All dimensions are shown in millimeters 8 Chapter 8 Specifications 223 ...