Status Reporting Commands
SYSTem:ERRor?
Query the multimeter’s error queue. Up to 20 errors can be stored in the
queue. Errors are retrieved in first-in-first out (
FIFO
) order. Each error
string may contain up to 80 characters.
STATus:QUEStionable:ENABle <
enable value
>
Enable bits in the Questionable Data enable register. The selected bits are then
reported to the Status Byte.
STATus:QUEStionable:ENABle?
Query the Questionable Data enable register. The multimeter returns a
binary-weighted decimal representing the bits set in the enable register.
STATus:QUEStionable:EVENt?
Query the Questionable Data event register. The multimeter returns a
decimal value which corresponds to the binary-weighted sum of all bits
set in the register.
STATus:PRESet
Clear all bits in the Questionable Data enable register.
*CLS
Clear the Status Byte summary register and all event registers.
*ESE <
enable value
>
Enable bits in the Standard Event enable register. The selected bits are then
reported to the Status Byte.
*ESE?
Query the Standard Event enable register. The multimeter returns a
decimal value which corresponds to the binary-weighted sum of all bits
set in the register.
Chapter 4 Remote Interface Reference
Status Reporting Commands
144
Содержание 34401A
Страница 12: ...1 Quick Start 1 ...
Страница 26: ...2 Front Panel Menu Operation 2 ...
Страница 50: ...3 Features and Functions 3 ...
Страница 103: ...4 Remote Interface Reference 4 ...
Страница 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Страница 170: ...5 Error Messages 5 ...
Страница 182: ...6 Application Programs 6 ...
Страница 196: ...7 Measurement Tutorial 7 ...
Страница 214: ...8 Specifications 8 ...
Страница 222: ...Product Dimensions TOP Product Dimensions All dimensions are shown in millimeters 8 Chapter 8 Specifications 223 ...