2
Set up the Format menu.
a
Touch Configuration, then select Format. In the Format menu, touch Timing
Acquisition Mode, then select 1M (2M) Sample Full Channel 250 MHz.
b
Touch the field to the right of the Pod 1 field, then select ECL.
c
Touch the field showing the channel assignments for Pod 1. Deactivate all channels
by selecting Clear. Using the knob, move the cursor to Channel 0. Touch the asterisk
field to put an asterisk in the channel position, activating the channel, then touch
Done.
3
Touch Format, then select Trigger. In the Trigger menu, touch Modify Trigger, then
select Clear Trigger, then select All.
Testing Performance
To test the time interval accuracy
3–53