Acquire the data
1
Enable the pulse generator channel 1 output (LED off).
2
In the logic analyzer Waveform menu, touch Run-Repetitive. Allow the logic
analyzer to acquire data for at least 100 valid runs as indicated in the pattern
statistics field.
3
When the logic analyzer has acquired more than 100 valid runs, touch Stop. The
Min X-O field in the logic analyzer Pattern Statistics menu should read
94.99–95.00
µ
s. The Max X-O field should read 95.00
–
95.01
µ
s. The Avg X-O field
should read 95.00
µ
s. Record the results in the performance test record.
Testing Performance
To test the time interval accuracy