To perform the self-tests and make the test connectors
The self-tests verify the correct operation of the logic analyzer module. Self-tests can
be performed all at once or one at a time. While testing the performance of the
module, run the self-test all at once.
The test connectors connect the analyzer to the test equipment.
Perform the self-tests
1
Disconnect all inputs.
2
In the System Configuration menu, touch Configuration. In the pop-up, touch Test.
3
Touch the box labeled Touch Box to Load Test System.
4
On the test system screen, touch Test System. Select the module to be tested. For
an HP 16555A, select 1M Sample LA as the module to be tested. For an HP 16555D,
select 2M Sample LA.
3–3
Содержание 16555A
Страница 4: ...The HP 16555A D Logic Analyzer iii ...
Страница 15: ...1 8 ...
Страница 97: ...3 70 ...
Страница 98: ...4 Calibrating ...
Страница 102: ...Troubleshooting Flowchart 1 Troubleshooting To use the flowcharts 5 3 ...
Страница 103: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 5 4 ...
Страница 104: ...Troubleshooting Flowchart 3 Troubleshooting To use the flowcharts 5 5 ...
Страница 105: ...Troubleshooting Flowchart 4 3 Troubleshooting To use the flowcharts 5 6 ...
Страница 117: ...5 18 ...
Страница 125: ...6 8 ...
Страница 126: ...7 Replaceable Parts Ordering 7 2 Replaceable Parts List 7 3 Exploded View 7 5 Replaceable Parts ...
Страница 130: ...Exploded View Exploded view of the HP 16555A D logic analyzer Replaceable Parts Exploded View 7 5 ...
Страница 131: ...7 6 ...
Страница 132: ...8 Block Level Theory 8 2 Self Tests Description 8 6 Theory of Operation ...