Preparing For Use
This chapter gives you instructions for preparing the logic analyzer module for use.
Power Requirements
All power supplies required for operating the logic analyzer are supplied through the
backplane connector in the mainframe.
Operating Environment
The operating environment is listed in chapter 1. Note the non-condensing humidity
limitation. Condensation within the instrument can cause poor operation or
malfunction. Provide protection against internal condensation.
The logic analyzer module will operate at all specifications within the temperature
and humidity range given in chapter 1. However, reliability is enhanced when
operating the module within the following ranges:
•
Temperature: +20
°
C to +35
°
C (+68
°
F to +95
°
F)
•
Humidity: 20% to 80% non-condensing
Storage
Store or ship the logic analyzer in environments within the following limits:
•
Temperature: -40
°
C to +75
°
C
•
Humidity: Up to 90% at 65
°
C
•
Altitude: Up to 15,300 meters (50,000 feet)
Protect the module from temperature extremes which cause condensation on the
instrument.
To inspect the module
1
Inspect the shipping container for damage.
If the shipping container or cushioning material is damaged, keep them until you have
checked the contents of the shipment and checked the instrument mechanically and
electrically.
2
Check the supplied accessories.
Accessories supplied with the module are listed in chapter 1, "Accessories Supplied."
3
Inspect the product for physical damage.
Check the module and the supplied accessories for obvious physical or mechanical defects. If
you find any defects, contact your nearest Hewlett-Packard Sales Office. Arrangements for
repair or replacement are made, at Hewlett-Packard’s option, without waiting for a claim
settlement.
2–2
Содержание 16555A
Страница 4: ...The HP 16555A D Logic Analyzer iii ...
Страница 15: ...1 8 ...
Страница 97: ...3 70 ...
Страница 98: ...4 Calibrating ...
Страница 102: ...Troubleshooting Flowchart 1 Troubleshooting To use the flowcharts 5 3 ...
Страница 103: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 5 4 ...
Страница 104: ...Troubleshooting Flowchart 3 Troubleshooting To use the flowcharts 5 5 ...
Страница 105: ...Troubleshooting Flowchart 4 3 Troubleshooting To use the flowcharts 5 6 ...
Страница 117: ...5 18 ...
Страница 125: ...6 8 ...
Страница 126: ...7 Replaceable Parts Ordering 7 2 Replaceable Parts List 7 3 Exploded View 7 5 Replaceable Parts ...
Страница 130: ...Exploded View Exploded view of the HP 16555A D logic analyzer Replaceable Parts Exploded View 7 5 ...
Страница 131: ...7 6 ...
Страница 132: ...8 Block Level Theory 8 2 Self Tests Description 8 6 Theory of Operation ...