5 Functions
65
5.8.4
Calibration Report Check
The calibration report can be checked here.
1.
Press the CAL. REPORT] on the DATA screen.
The SELECT COMP. screen appears.
2.
Select the measurement component to be checked.
The CAL. REPORT screen appears.
Up to 20 calibration dates and zero/span factors are displayed by measurement
component and by zero/span in chronological order from the latest date.
The screens show examples for UV.
Reference
About the Calibration coefficient" (page 15)
Tip
[ZERO]/[SPAN]: Switches between ZERO CAL. REPORT and SPAN CAL. REPORT.
[
], [
]: Switches to the previous and next data.
[DELETE]: Deletes all of the calibration report.
When changing the measurement component to display, press
[
]
to go back to the SELECT
COMP. screen.
[DATA]
MEAS. screen
Refer to "
Data Screen Display" (page 60)
DATA screen
[CAL.
REPORT]
[UV]