8 Troubleshooting
95
8
Troubleshooting
8.1
Table of Status and Operation
8.2
Table of alarm and Operation
Item
Status
LED
(POWER/ALARM)
◎
: green light on
○
: green blinking
Contact
(batch
alarm)
Contact
(maintenan
ce)
Output status
Alarm
display
Alarm
history
Common
MEAS.
◎
Instantaneous
value output
MAINTENANCE
○
Maintenance output
○
Maintena
nce
CALIBLATION
○
Maintenance output
○
Maintena
nce
DATA
◎
Instantaneous
value output
Item
Status
LED
(POWER/ALARM)
◎
: green light on
●
: red light on
○
: green blinking
△
: red blinking
Contact
(batch
alarm)
Contact
(analyzer
alarm)
Output status
Alarm
display
Alarm
history
Common
WARM-UP
○
Warm-up output
○
○
CELL MOTOR
ERR.
●
○
Alarm output
○
○
SENSOR
COMM.ERR.
*1
●
○
○
Alarm output
○
○
SAMPLE LACK
△
Instantaneous
value output
○
○
PIO COMM.ERR.
*2
●
○
○
Alarm output
○
○
POWER ON/OFF
Instantaneous
value output
○
BATTERY ERR.
△
Instantaneous
value output
○
○
LAMP ERR.
●
○
Alarm output
○
○
UV
ZERO CAL.
△
Instantaneous
value output
○
○
SPAN CAL.
△
Instantaneous
value output
○
○
VIS
ZERO CAL.
△
Instantaneous
value output
○
○
SPAN CAL.
△
Instantaneous
value output
○
○
COD
HIGH
△
Instantaneous
value output
○
○
LOW
△
Instantaneous
value output
○
○