9 Reference Data
103
Contact output
Point: 6
Type: arbitrary 4 items from POWER OFF, MAINTENANCE (standard, fixed
allocation)
BATCH ALARM, COD MAX ALARM, TURB MAX ALARM (standard), LIGHT
SOURCE MALFUNCTION, NO SAMPLE, CLEANING MOTOR ERROR,
ANALYZER ERROR (selectable)
Content: POWER OFF: when the power OFF occurred.
BATCH ALARM: CLEANING MOTOR ERROR, LIGHT SOURCE
MALFUNCTION, ANALYZER ERROR
MAINTENANCE: at in-maintenance mode and calibration mode, and when
maintenance switch is ON.
Specification: contact output without voltage, a contact
Contact rating: 125 V AC/0.3 A, 30 V DC/1 A
(however, for load resistance) each output COM-independent
Contact input
Point: 2
Type: Float switch input, time adjustment input
Specification: contact input without voltage (open collector can be connected),
insulated input
ON resistance: Max. 100
, Open voltage: 5.5 V DC, Short circuit current:
Max. 10 mA
Communication
Interface: RS-232C compliant
Communication speed: 19200 bps
Data memory
Memorize the data of measurement value of the measurement item in the
main body.
Memorized data can be transferred to the CF card.
Memory interval: 1 minute or 1 hour
Memory time: on the hour every hour
Data memory time: interval of 1 minute: approximately 10 days
interval of 1 hour: approximately 1 year
* The latest data is memorized.
Light source/detector
Light source: low pressure mercury-vapor lamp, detector: silicon photo diode
Piping sockets
Sample inlet Rp-1/2 female
Bypass outlet Rc-1/2 female
Overflow outlet (1) Elbow fitting, 13A (nominal)
Overflow outlet (2) Elbow fitting, 20A (nominal)
Drain outlet Rc-1/2 female
Discharge Tube coupling, 50A (nominal)
Structure
Weather-proofed for outdoor installation (equivalent to IP54)
Main material of each unit
SUS, PVC, PP, CR, SiO
2
Power
100 V to 230 V AC ±10%, 50/60 Hz
Power consumption
100 V to 120 V AC, Max. 45 VA
200 V to 230 V AC, Max. 60 VA
Mass
Control unit: approximately 5.0 kg
Analyzer unit: approximately 5.6 kg
Dimension
Control unit: 240 (W)
×
104 (D)
×
320 (H) mm
Analyzer unit: 200 (W)
×
180 (D)
×
403 (H) mm
(Protruding portion are not included.)
Painting color
Control unit: Munsel 5PB 8/1
Conditions for installation
Flat and stable places free from strong vibration or shock
Places free from dirt or dust, or where no corrosive gases will be
generated.
At the atmospheric pressure
Places not subject to direct sunlight
Places well-ventilated
Altitude of 2000 m or below