4 Calibration
34
8.
Enter the concentration value of compounded TURB calibration solution with
numeric keypad, and press [SET],
Input range: 0.00 to 9999.99
The screen returns to the TURB. CAL. screen.
Note
The decimal place of input range varies according to the setting in "5.5.2 Setting of Decimal Place for
each Component" (page 48).
Note
Enter the concentration value of compounded calibration solution for TURB calibration value.
Tip
[CLEAR]: Clears all input numeric value.
[BACK]: Clears the rightmost number.
9.
After waiting for 3 minutes and confirming that the measurement value is
stabilized, press [SPAN].
The “Start span cal.?” screen is displayed.
10.
Press [YES] to start the span calibration.
“Calibrating” screen is displayed.
After calibration, “Calibration finished” screen is displayed.
11.
Press [YES].
The span calibration value for TURB (turbidity) is updated.
12.
The TURB span calibration is finished.
Set the analyzer unit in the measuring tank.
Note
Do not reuse the calibration solution used in the calibration tank. The accuracy of calibrated value
may be lost.