5 Functions
38
2.
Output signal is allocated.
Reference
"6 External Input and Output" (page 72)
Setting item
Description
Initial setting
ANALOG OUTPUT A1 Sets to output either one of the following
measurement component to the analog output
terminal from A1 to A3.
Setting range:
UV, VIS, UV-VIS, COD, TURB, NONE
COD
ANALOG OUTPUT A2
TURB
ANALOG OUTPUT A3
UV-VIS
SIGNAL OUTPUT 01
Sets to output either one of the following items to the
output terminal of contact output from 01 to 04.
Setting range:
TOTAL ALARM, MAX COD, MAX TURB, MIN COD,
MIN TURB, LAMP ERR., NO SAMPLE, MOTOR
ERR., ANALYZE ERR., WARM-UP
TOTAL ALARM
SIGNAL OUTPUT 02
MAX COD
SIGNAL OUTPUT 03
MAX TURB
SIGNAL OUTPUT 04
NO SAMPLE
SIGNAL OUTPUT 05
Outputs the following item to the output terminal 05 of
contact output.
POWER OFF (cannot be changed)
POWER OFF
(fixed)
SIGNAL OUTPUT 06
Outputs the following item to the output terminal 06 of
contact output.
MAINTENANCE (cannot be changed)
MAINTENANCE
(fixed)