General Setup
42
● P5.1 Auto Stable, Auto Hold and Real Time Mode Setup
Auto Stable (AS) Mode
-
The meter shows live readings; annunciator blinks until
reading is stable.
Auto Hold (AH) Mode
-
The meter locks the stable reading; annunciator blinks until
reading is stable and then lights up.
Real Time (RT) Mode
-
The meter shows live readings; both and annunciators are
inactivate.
1. Press the
key,
P1 COND
screen appears.
2. Press the
key,
P2 TDS
screen appears.
3. Press the
key,
P3 SAL
screen appears.
4. Press the
key,
P4 DATA
screen appears.
5. Press the
key,
P5 GEN
screen appears.
6. Press the
ENT
key,
P5.1 STBL
screen appears.
7. Press the
ENT
key, the default stability mode is
AS
(auto stable).
8. Use the
keys to change the stability mode as
AH
(auto hold) or
RT
(real time).
9. Press the
ENT
key,
P5.1 STBL
screen appears. This indicates completion of the
stability mode selection.
key
key
key
key
key
key
key
Содержание LAQUA-EC1500
Страница 1: ...Instruction Manual Benchtop Conductivity Meter LAQUA EC1500...
Страница 2: ......
Страница 4: ...II MEMO...
Страница 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Страница 24: ...Product Overview 6...
Страница 46: ...Conductivity Setup 28...
Страница 50: ...TDS Setup 32...
Страница 64: ...General Setup 46...
Страница 76: ...Printout Summary 58 Appendix 2 Printer Format Measurement Conductivity Resistivity Salinity...
Страница 77: ...Printout Summary 59 TDS Printer Format Data Log...
Страница 78: ...Printout Summary 60 Printer Format Calibration Conductivity...
Страница 79: ......