![horiba LAQUA-EC1500 Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/horiba/laqua-ec1500/laqua-ec1500_instruction-manual_2162249054.webp)
Salinity Setup
36
● P3.3 Erase Calibration Data
1. Press the
key,
P1 COND
screen appears.
2. Press the
key,
P2 TDS
screen appears.
3. Press the
key,
P3 SAL
screen appears.
4. Press the
ENT
key,
P3.1 UNIT
screen appears.
5. Press the
key,
P3.2 TYPE
screen appears.
6. Press the
key,
P3.3 C.CLr
screen appears.
7. Press the
ENT
key,
C.CLr NO
appears with
NO
as default set up.
8. Use the
keys to change the set up to
YES
. This erases the calibration data.
9. Press the
ENT
key,
P3.3 C.CLR
screen appears. This indicates erasure of
calibration data.
key
key
key
key
key
key
key
Содержание LAQUA-EC1500
Страница 1: ...Instruction Manual Benchtop Conductivity Meter LAQUA EC1500...
Страница 2: ......
Страница 4: ...II MEMO...
Страница 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Страница 24: ...Product Overview 6...
Страница 46: ...Conductivity Setup 28...
Страница 50: ...TDS Setup 32...
Страница 64: ...General Setup 46...
Страница 76: ...Printout Summary 58 Appendix 2 Printer Format Measurement Conductivity Resistivity Salinity...
Страница 77: ...Printout Summary 59 TDS Printer Format Data Log...
Страница 78: ...Printout Summary 60 Printer Format Calibration Conductivity...
Страница 79: ......