Conductivity Calibration
12
Auto Conductivity Calibration
1. After placing the conductivity electrode in the standard solution, press the
key.
2. Meter displays “Auto cal” as per set calibration method and starts measuring various
calibration values with a
blinking on screen.
3. Wait for the
to stabilize (stable calibration reading). When it stabilizes, there will be a
sound.
4. Press the
ENT
key to confirm and save calibration data.
5. Meter displays
DONE
indicating end of the conductivity calibration procedure.
6. Repeat for other calibration points as required.
7. You can calibrate one point for each range.
To view and /or print calibration data, press the
key when you are in the
CAL
mode. The display scrolls through the calibrated values and calibration factor (average
calibration factor for multi-point calibration).
key
key
Note
Содержание LAQUA-EC1500
Страница 1: ...Instruction Manual Benchtop Conductivity Meter LAQUA EC1500...
Страница 2: ......
Страница 4: ...II MEMO...
Страница 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Страница 24: ...Product Overview 6...
Страница 46: ...Conductivity Setup 28...
Страница 50: ...TDS Setup 32...
Страница 64: ...General Setup 46...
Страница 76: ...Printout Summary 58 Appendix 2 Printer Format Measurement Conductivity Resistivity Salinity...
Страница 77: ...Printout Summary 59 TDS Printer Format Data Log...
Страница 78: ...Printout Summary 60 Printer Format Calibration Conductivity...
Страница 79: ......