Salinity Calibration
16
Calibration
1. After placing the conductivity electrode in the standard solution, press the key.
2. Meter starts measuring various calibration values with a
blinking on screen.
3. Wait for the
to stabilize (stable calibration reading). When it stabilizes, there will be a
sound.
4. Use the
keys to adjust the salinity value.
5. Press the
ENT
key to confirm and save calibration data.
6. Meter displays
DONE
indicating end of the salinity calibration procedure.