![horiba LAQUA-EC1500 Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/horiba/laqua-ec1500/laqua-ec1500_instruction-manual_2162249033.webp)
Salinity Calibration
15
● Salinity Calibration
Calibration is necessary for accurate salinity measurement. To perform salinity calibration,
follow the procedure detailed below:
Prerequisites
• Clean the conductivity electrode with DI (deionized) water and wipe it with tissue paper.
• Switch on the EC meter and plug in the conductivity electrode.
• Prepare standard solution required for calibration.
• Press the
key to keep the meter in
SAL
mode.
• Dip the conductivity electrode in the standard solution till the hole at the upper part of the
electrode is immersed.
• Before salinity calibration, set the required salinity method. In LAQUA-EC1500 benchtop
meter, available salinity methods are;
-NACL
-SEA.W
(Sea water)
• To set a desired salinity method, refer “P3.2 Salinity Type Setup” on page 35.
• User can adjust the salinity value by calibration.
To abort an ongoing calibration process at any point of time, press the
key.
Note
Tip
Содержание LAQUA-EC1500
Страница 1: ...Instruction Manual Benchtop Conductivity Meter LAQUA EC1500...
Страница 2: ......
Страница 4: ...II MEMO...
Страница 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Страница 24: ...Product Overview 6...
Страница 46: ...Conductivity Setup 28...
Страница 50: ...TDS Setup 32...
Страница 64: ...General Setup 46...
Страница 76: ...Printout Summary 58 Appendix 2 Printer Format Measurement Conductivity Resistivity Salinity...
Страница 77: ...Printout Summary 59 TDS Printer Format Data Log...
Страница 78: ...Printout Summary 60 Printer Format Calibration Conductivity...
Страница 79: ......