Conductivity Setup
24
● P1.3 Calibration Mode Setup
1. Press the
key,
P1 COND
screen appears.
2. Press the
ENT
key,
P1.1 CELL
screen appears.
3. Press the
key,
P1.2 UNIT
screen appears.
4. Press the
key,
P1.3 A.CAL
appears.
5. Press the
ENT
key,
A.CAL ON
screen appears with
ON
as default set up.
6. Use the
keys to change the set up to
OFF
. This enables the manual calibration
mode.
7. Press the
ENT
key,
P1.3 A.CAL
screen appears. This indicates completion of auto
calibration mode setup.