Conductivity Setup
27
● P1.6 Erase Calibration Data
1. Press the
key,
P1 COND
screen appears.
2. Press the
ENT
key,
P1.1 CELL
screen appears.
3. Press the
key,
P1.2 UNIT
screen appears.
4. Press the
key,
P1.3 A.CAL
appears.
5. Press the
key,
P1.4 T.CFF
appears.
6. Press the
key,
P1.5 T.rEF
appears.
7. Press the
key,
P1.6 C.CLr
appears.
8. Press the
ENT
key,
C.CLr NO
screen appears with
NO
as default set up.
9. Use the
keys to change the set up to
YES
. This erases the calibration data.
10. Press the
ENT
key,
P1.6 C.CLr
screen appears. This indicates erasure of
calibration data.
key
key
key
key
key
key