4 Maintenance and Troubleshooting
4.2 Troubleshooting
D-51
63
●
The indicated value fluctuates
When there is a problem with the electrode...
When there is a problem with the main unit of the
pH meter...
Cause
How to solve problem
The responsive
membrane is
dry or dirty.
Wash the responsive membrane.
The responsive
membrane is
damaged or
worn out.
Replace the electrode.
There are air
bubbles on the
electrode.
Shake the electrode to remove the
air bubbles.
There is no
reference
solution
remaining.
Fill the electrode with new reference
solution, as noted in the electrode
operation manual.
The wrong
reference
solution is being
used.
Use the correct reference solution.
Cause
How to solve problem
There is a motor
or other device
causing
electrical
interference.
Move the meter to a place where it
is not subject to dielectric effects.
Be sure to ground devices that are
using commercial electricity.
The electrode is
not connected
correctly.
Connect the electrode correctly.