3 Functions
3.2 pH calibration history display
32
HORIBA
3.2
pH calibration history display
The latest calibration and repeatability check information
can be checked.
Refer to
pH repeatability check on page 26.
●
pH calibration history
1.
Press the CAL DATA key in the pH Measure-
ment mode.
2.
Pressing the CAL DATA key toggles between
the latest calibration date and the latest check
date.
The latest check date is not displayed if no repeatability
check has been performed.
Latest calibration date
Latest check date