4 Maintenance and Troubleshooting
4.1 pH electrode maintenance
D-51
51
4
Maintenance and Troubleshooting
This chapter explains how to perform daily meter
maintenance and how to deal with error messages.
Daily maintenance is vital in assuring accurate
measurement and preventing breakdowns before they
occur. Maintenance of the electrodes is especially
important; if ignored, various problems and erroneous
measurements may result. This meter is equipped with
a convenient error message function. If an error
message is displayed, be sure to take appropriate
action.
4.1
pH electrode maintenance
Maintain your electrodes by referring to the following
information or to the operation manuals for the
electrodes.
●
Maintenance after daily use
After taking measurements, wash the electrode using
pure water (de-ionized water), wipe off the water from
the electrode with filter paper or tissue paper, and store it
with its cap on.
Injury warning
Glass fragments can cause injury.
The outer tube of the electrode and the
tip of the electrode are made of glass.
Use care not to break them.
Caution