4 Maintenance and Troubleshooting
4.2 Troubleshooting
D-51
59
●
ERR No.05 Electrode sensitivity error (pH)
Explanation
The electrode sensitivity is either 105% or more or 85%
or less than the theoretical sensitivity.
Cause
How to solve problem
The electrode is
dirty.
Wash the electrode.
The electrode is
cracked.
Replace the electrode.
Calibration was
not performed
correctly.
Redo the calibration correctly.
There is a
problem with the
standard solution.
Use fresh standard solution.
The electrode is
not connected
correctly.
Connect the electrode correctly.
Electrode is not
submerged
deeply enough to
cover reference
junction.
Immerse the electrode in the
sample at least three centimeters
deep.