4 Maintenance and Troubleshooting
4.2 Troubleshooting
58
HORIBA
●
ERR No.04 Asymmetric potential error
Explanation
The asymmetric potential of the electrode is 45 mV or
more.
Cause
How to solve problem
The electrode is
dirty.
Wash the electrode.
The electrode is
cracked.
Replace the electrode.
The reference
solution
concentration is
fluctuating.
Replace the internal solution in the
reference electrode.
The electrode is
not connected
correctly.
Connect the electrode correctly.
The electrode is
not submerged
deeply enough to
cover the liquid
junction (tip).
Immerse the electrode in the
sample at least three centimeters
deep.
There is problem
with the standard
solution.
Prepare new standard solution.