Error messages and trouble shooting
58
●
Trouble shooting
T
his section describes causes and actions to take for problems that customers frequently ask.
The indicated value fluctuates
< Problem with the electrode >
<
Problem with the instrument
>
< Problem with the sample >
Cause
How to solve problem
The electrode is dirty.
Clean the electrode.
The electrode is broken.
Replace the electrode.
The level of reference electrolyte gel.
Replace the electrode.
Cause
How to solve problem
There is a motor or other device
causing electrical interference.
Measure at a place where no influence from induction
is given. Ground all AC-powered equipment.
The electrode is not connected
correctly.
Connect the electrode properly.
Cause
How to solve problem
Electrode is not immersed enough to
cover liquid junction.
The electrode must be immersed up to the liquid
junction.
As a guide, immerse to at least 3 cm from the tip of
the electrode.
The stability of electrode is affected by
the sample solution.
It is important to select an electrode that is
appropriate for the sample. Consult your dealer. To
confirm an electrode that is appropriate for the
sample, check the pH electrode selection guide in our
catalogue, or refer to our website.
Содержание LAQUA-PD210
Страница 1: ......
Страница 2: ......
Страница 49: ...Data set up 35 DO mode...
Страница 51: ...Data set up 37 DO mode...
Страница 53: ...Data set up 39 DO mode...
Страница 56: ...General set up 42 DO mode...
Страница 58: ...General set up 44 DO mode...
Страница 60: ...General set up 46 DO mode...
Страница 62: ...General set up 48 DO mode...
Страница 65: ...Clock set up 51 DO mode...
Страница 67: ...Clock set up 53 DO mode...
Страница 86: ...Specifications 67...
Страница 88: ......