DO calibration
19
●
1
st
point calibration
1. After placing the DO electrode in the high concentration solution, press the
key.
Meter starts checking various calibration values with a blinking
on screen.
2. Wait for the
to stabilize (stable calibration reading).
3. Use the
keys to adjust the DO reading.
4. Press the
ENT
key to confirm and save calibration data.
5. Meter displays
CAL
DONE
indicating end of the calibration procedure.