pH set up
27
●
P 1.2 Calibration alarm setup
1. Press the
key,
P1 PH
screen appears.
2. Press the
ENT
key,
P1.1 BUFF
screen appears.
3. Press the
key,
P1.2 C.ALr
screen appears.
4. Press the
ENT
key, by default
DAYS ----
appears.
5. Use the
keys to adjust the calibration alarm interval for next calibration.
6. Press the
ENT
key,
P1.2 C.ALr
screen appears. This indicates completion of
calibration alarm setup.