pH calibration
11
■
Calibration
This section describes the basic calibration method using LAQUA PD200 handheld meter, pH
electrode and DO electrode.
●
pH Calibration
Calibration is necessary for accurate pH measurement. To perform pH calibration, follow the
procedure detailed below.
Prerequisites
• Clean the pH electrode with DI (deionized) water and wipe it with tissue paper.
• Switch on the PD meter and plug in the pH electrode.
• Prepare buffer solution required for calibration.
• Keep the meter in pH measurement mode.
• Dip the pH electrode at least 3 cm in the buffer solution.
• Perform two-point calibration using:
pH 7 and 4 for acidic sample.
pH 7 and 10 for alkaline sample.
• Perform three-point calibration using pH7, 4 and 10 if you are unsure of the expected
sample pH value. It is recommended to calibrate with pH7 first.
• Default buffer setup is
BUFF USA
. If you want to change to
BUFF NIST
or
BUFF DIN
,
refer to “P 1.1 Buffer selection” on page 26.
• To abort an ongoing calibration process at any point of time, press the
key.
• It is recommended to clear the previous calibration data before performing calibration. For
erasing the calibration data, refer to “P 1.3 Erase calibration data” on page 28.
Note
Tip
Содержание LAQUA-PD210
Страница 1: ......
Страница 2: ......
Страница 49: ...Data set up 35 DO mode...
Страница 51: ...Data set up 37 DO mode...
Страница 53: ...Data set up 39 DO mode...
Страница 56: ...General set up 42 DO mode...
Страница 58: ...General set up 44 DO mode...
Страница 60: ...General set up 46 DO mode...
Страница 62: ...General set up 48 DO mode...
Страница 65: ...Clock set up 51 DO mode...
Страница 67: ...Clock set up 53 DO mode...
Страница 86: ...Specifications 67...
Страница 88: ......