General set up
47
●
P 3.4 Reset meter (factory default)
1. Press the
key,
P1 PH/DO
screen appears.
2. Press the
key,
P2 DATA
screen appears.
3. Press the
key,
P3 GEN
screen appears.
4. Press the
ENT
key,
P3.1 STBL
screen appears.
5. Press the
key,
P3.2 A.OFF
screen appears.
6. Press the
key,
P3.3 °C°F
screen appears.
7. Press the
key,
P3.4 r.SET
screen appears.
8. Press the
ENT
key, default meter re-setup is
NO
.
9. Use the
key to set it
YES
.
10. Press the
key. Meter displays
DONE
and automatically switches off.
pH mode
Содержание LAQUA-PD210