DO set up
31
●
P1.2 Barometric pressure setup
1. Press the
key,
P1 DO
screen appears.
2. Press the
ENT
key,
P1.1 SAL
screen appears.
3. Press the
key,
P1.2 PRES
screen appears.
4. Press the
ENT
key, by default
PRES 101.3
appears.
5. Use the
keys to adjust the barometric pressure in between 10.0 to 200.0 kPa.
6. Press the
ENT
key,
P1.2 PRES
screen appears. This indicates completion of
barometric pressure setup.