DO calibration
16
●
Air calibration
1. Hold the DO electrode still in clean air.
2. Press the
key. Meter starts checking various calibration values with a blinking
on
screen.
3. Wait for the
to stabilize (stable calibration reading).
4. Press the
ENT
key to confirm and save calibration data.
5. Meter displays
DONE
indicating end of the air calibration procedure.
The electrode responds slightly differently to atmospheric air as compared to water. As such,
100% calibration in air will be shown as approximately 105%. Don't be alarmed by this.The
reading in water will be accurate.