DO set up
29
●
P1 DO setup
Using P1 DO setup function of the meter, you can:
• Set salinity value
• Set barometric pressure
• Erase calibration data
To set the DO functions using LAQUA PD200 handheld meter, follow the procedure detailed
below:
Prerequisites
Switch on the PD meter.
• Default salinity value is
0.0 ppt
. You can set a value in between 0.0 to 40.0 ppt.
• Default barometric pressure is
101.3 kPa
. You can set a value in between 10.0 to 200.0
kPa.
• Erasing previous calibration data is recommended for accurate calibration. Default setup is
NO
but to erase the calibration data, you need to change the setup to
YES
.
To return to the measurement mode, press the
key.
Note
Tip