Page 84
STT700 Series HART/DE Option User’s Manual
Revision 1
11.
Select
OK
to acknowledge.
7.1.4 Correct Input at Sensor 1 / 2 Calibration High Point
1.
Select the S 1 / 2 Cal Hi Corrects button. This message appears.
1.
Check the device is not in write protect mode and press OK
2.
Enter calibration date (MM/DD/YY), OK
3.
Enter calibration Time (Hour), OK. Entre calibration time (minute), OK
4.
Apply input on Sensor 1.