Revision 1
STT700 Series HART/DE Option User’s Manual
Page 75
a.
Select
Set Output To 0%
or
100%
. You will be prompted to confirm that you want
to place the transmitter in output mode.
b.
Verify that the loop is in manual control. In output mode, output current is fixed at the
0% or 100% level as selected in the TRIM DAC CURRENT box in the previous step.
c.
Select
Yes
, and observe the loop current level. A meter reading of 4 mA corresponds
to 1 volt.
d.
Use the MC Toolkit to adjust the loop current to the Zero Percent level (4mA). If the
current is low, tap the
Increment
button; if the current is high, tap the
Decrement
button. Note that the value on the meter changes accordingly. If the error is large,
accelerate the adjustment rate by changing the Step Size to 10 or 100.
e.
After establishing the zero current level (4 mA), select
Set Output To 100%
. A meter
reading of 20 mA corresponds to 5 volts.
f.
Use the
Increment
or
Decrement
button, as necessary to adjust the output current to
20 mA. When the current reaches the 20 mA level, select
Clear Output
; the button
will change to half-intensity.
5.
Change the display in output mode as follows:
a.
Selecting the
Back
button before selecting the
Clear Output
button, you will be
prompted to confirm that you want to clear the output.
b.
If you want to stay in output mode while viewing other displays, select
Yes
;
otherwise, select
No
and the
Clear Output
button.