Satellite XT R Version
5-4
Manual K-Factor
This submenu is used to manually calibrate the
instrument by entering a new K-factor. This factor is
a multiplication or correction value used to calibrate
the response of the instrument to a specific gas
concentration. The factory default K-Factor for all
gases is 1.00. Use the <set> key to enter. The
instrument will display the actual Kfactor.
< Actual K-Factor>
Use the cursor keys <left> or <right> to reach the
desired position. Use the cursor keys <up> and
<down> to select the figures you wish to enter. The
K-factor must be in the range of 0.20 and 5.00.
After a new value has been entered, actuate the <set>
key to confirm or <esc> to exit.
! Note:
Individual settings for the manual K-factor will automatically revert to the default value 1.00,
whenever the sensor is replaced by a new sensor.
p l e a s e e n t e r !
M A N U A L K - F A C T O R
1
. 0 0
s e l e c t :
2
o f
2
C A L I B R A T I O N
MANUA L K - F AC TOR
Calibration - Section 5