Satellite XT R Version
7-2
7.1.2 Faults
Instrument faults refer to a problem which prevents the instrument from operating properly
and interferes with the ability to monitor or document concentration alarms.
If the instrument has detected an instrument fault, the green status LED is off and the fault
relay is activated. The LCD display is flashing while indicating the specific fault message.
Fault - ADC Not Ready !
This indicates that there is a problem with the internal
electronics. The AD Converter has failed.
Switch power supply for this unit off and on again. If the
instrument still shows this message, switch off power supply
for this unit and contact MST for further instructions.
Fault - No Sensor !
This message indicates that no sensor is inserted or that
a sensor type not suitable for the Satellite XT is used. To
eliminate the problem, insert the correct sensor.
Fault - Wrong Sensor !
This message indicates that the sensor installed is not the
sensor assigned to this particular instrument.
To correct the problem, insert the sensor assigned and
documented to this instrument. The data are stated in the test
certificate provided with the shipment.
Fault - Replace Sensor !
This message indicates that the sensor presently installed is
depleted (the warning CHECK SENSOR ! has been ignored).
To correct the problem, replace the old sensor immediately.
! This function is not available for instruments with
oxygen sensor.
- - - - F AU L T - - - -
ADC NOT RE ADY !
- - - - F AU L T - - - -
NO S ENSOR !
- - -
WR
-
ON
F A
G
U L
S E
T
NS
- -
OR
- -
!
- - - - F AU L T - - - -
RE P L ACE S ENSOR !
Troubleshooting - Section 7
Содержание Satellite XT
Страница 1: ...Guide to Operation Satellite XT R Version Including Options Extractive Module XT Pyrolyzer Module XT ...
Страница 15: ...Satellite XT R Version 2 8 Installation and Start Up Section 2 ...
Страница 49: ...Satellite XT R Version 6 16 Configuration Section 6 ...
Страница 53: ...Satellite XT R Version 7 4 Troubleshooting Section 7 ...