9-6
9.8
Technical Specifications
Power Requirements
voltage
power is provided by the Satellite XT,
power supply range 18 - 24 VDC
consumption
max. 2.4 W
Physical Dimensions
size
78 x 95,5 x 50 mm
(W x H x D)
3,1" x 3,8" x 2,0"
(W x H x D)
weight
500 g
17.6 oz
Housing Protection Class
IP 30
RFI / EMC
EN 50270
Operating Conditions
sample tubing
OD 1/4” ID 3/16” PTFE or PFA
response time
< 30 sec. with a tubing length of 10m
temperature
0 °C ... +40 °C
+32 °F ... +104 °F
pressure
700 ... 1300 hPа
humidity
Maximum negative pressure
(at pump)
Maximum sample line length
Recommended sample
line length
20 ... 90 % r.h.
-150mbar
50 m (see note in section 9.6)
10 m
Part Number
version for toxic/corrosive gases
(except C
5
F
8
and SF
6
)
20404-0200
version for combustible gases
20404-0250
version for C
5
F
8
and SF
6
20404-0300
Warning:
Extractive Module 20404-0200 must only be used with Pyrolyzer
Modules 20408-0110, 20408-0112, 20408-0116 and 20408-0122.
Extractive Module 20404-0300 must only be used with Pyrolyzer
Modules 20408-0114 and 20408-0120.
Extractive Module XT - Section 9
Содержание Satellite XT
Страница 1: ...Guide to Operation Satellite XT R Version Including Options Extractive Module XT Pyrolyzer Module XT ...
Страница 15: ...Satellite XT R Version 2 8 Installation and Start Up Section 2 ...
Страница 49: ...Satellite XT R Version 6 16 Configuration Section 6 ...
Страница 53: ...Satellite XT R Version 7 4 Troubleshooting Section 7 ...