2.4 Wiring
There are two permanently attached cables extending
from the bottom of the instrument, a 2-wire shielded
cable with a length of 2 meters and a 6-
wire shielded cable with a length of 3 meters.
The 2-wire shielded cable is required to power the
instrument. The open end of the cable should be
terminated in a junction box.
The 6-wire shielded cable provides a pair of
connections for each of the 3 internal relays for
activation of visual or audible external alarm devices.
Power Supply Connections
P+
White
Power +
P–
Brown
Power -
S
Shield
cable shield
Relay Contacts
1
White
Alarm 1 Relay
1
Brown
Alarm 1 Relay
2
Green
Alarm 2 Relay
2
Yellow
Alarm 2 Relay
3
Grey
Fault Relay
3
Pink
Fault Relay
S
Shield
Cable shield
Note:
All wiring for the Satellite XT and its related equipment must be in compliance with any local
electrical and fire codes. The Satellite XT wiring must be kept away from any high power lines.
All shielding for the wiring must be connected together and grounded at only one point in the
system.
2-5
Installation and Start-Up - Section 2