INTELLIDOX DOCKING MODULE USER MANUAL || CONFIGURE SETTINGS VIA
SAFETY SUITE DEVICE CONFIGURATOR AND SAFETY SUITE SOFTWARE
HONEYWELL
PAGE 70 OF 119
Setting the Module Time
1. Start Safety Suite Device Configurator software, and then log in as an administrator.
2. Select
Configure Devices via IntelliDoX
on the
Devices
menu. The
IntelliDoX
Configuration Selection
dialog box is displayed.
3. Select a configuration file and click
OK
. The
IntelliDoX Device Configuration
dialog
box is displayed.
4. Select the tab for a compatible detector, and then select the
IntelliDoX
Configuration
tab.
5. Click on the
Time zone
selector. Select your time zone from drop down menu.
6. Verify that the
Current date and time
displayed under
Time zone settings
is
correct. The date is synchronized with the PC on which Safety Suite Device
Configurator software is installed. If the date is incorrect, change the date on the PC.
If the time is incorrect, click on the
Time zone
selector to select a different time zone.
7. When the correct date and time are displayed, click
Set IntelliDoX Time
. The
IntelliDoX Time Confirmation
dialog box is displayed.
8. Click
Save to IntelliDoX
. The
IntelliDoX Selection
dialog box is displayed. Select
one or more IntelliDoX modules, and then click
OK
.