INTELLIDOX DOCKING MODULE USER MANUAL || ASSEMBLE MODULES
HONEYWELL
PAGE 18 OF 119
Mount on a Wall
You can use the back panel to mount the module securely to a wall. Because wall materials
vary, mounting hardware is not provided. Honeywell strongly recommends that wall-mounting
is performed by a qualified installation contractor. Sufficient expertise is required to:
1. Determine the strength of the wall supporting the module; and
2. Select mounting hardware appropriate to the weight and normal use of the module.
Before you Begin
1. Select a suitable installation location. Each module or group of connected modules
requires:
•
Access to an appropriate power source.
•
Access to a network connection, if using.
•
Access to calibration gas sources.
•
Adequate clearance for the power cord, cables and tubing.
•
Adequate clearance for normal use, maintenance and cleaning.
2. Have ready appropriate mounting hardware. Each module requires sufficient hardware for
4 anchor points.
3. Have ready tools and accessories required to install the mounting hardware.
Incomplete or improper mounting may result in property damage and/or personal injury.
Do not install near heat sources or on vibrating surfaces.
The wall and the mounting hardware must be strong enough to support the equipment
during normal use.
The equipment must be used only in a normal atmosphere (20.9% v/v O
2
) that is free of
hazardous gas. Do not use the equipment in a hazardous area. Failure to adhere to this
caution can lead to fire and/or explosion.
The equipment is intended for indoor use only.
This equipment uses potentially harmful gas for calibrations. The equipment must be
attached to a venting system or be used in a well-ventilated area.